Verification of the definition of the HB11 beam in the INEL treatment planning software as applied to the NKI-head phantom
|
|
Auteur(s):
|
|
|
Gepubliceerd door:
|
Publicatie datum:
|
ECN
NUCLEAIR
|
1997
|
|
ECN publicatienummer:
|
Publicatie type:
|
ECN-C--97-102
|
ECN rapport
|
|
Aantal pagina's:
|
Volledige tekst:
|
21
|
Niet beschikbaar.
|
Samenvatting:
Results from the INEL Treatment Planning system were compared withmeasurements to verify the HB11 neutron beam of the High Flux Reactor (HFR,
Petten, Netherlands) in the treatment planning software. The calculations and
measurements were performed with the tissue-equivalent PMMA
(polymethylmetaacrylate) NKI-head phantom. The calculated thermal neutron
fluence rate (#phi#th), fast neutron dose rate (Dn), and gamma dose rate
(Dg) are compared with measurements. 1 fig., 3 tabs., 7 refs., 3 appendices